Drager Babylog 8000 Plus Ventilator
Drager Babylog 8000 Plus Ventilator
Product Specification
Model Number | Babylog 8000 Plus |
Usage/Application | ICU |
Brand | Drager |
Modes of ventilation | SIMV+VG,SIPPV+VG,PSV+VG |
Patient Age Group | Pediatric |
Operation Mode | Electricity Operated |
Inspiratory time | 0.1 to 2 seconds |
Expiratory Time | 0.2 to 30 seconds |
Weight | 32 lbs |
Power Source | Electric |
Inspiratory Pressure | 10 to 80 mbar |
Cpap | 0 to 25 mbar |
Size | 12.5 In (H) x 8.4 In (W) x 15.2 In (D) |
Base Flow | (VIVE) 1 to 30 L/min |
Condition | Refurbished |
Inspiratory Flow | 1 to 30 L/min |
Inspiratory Oxygen Concentration | 21 -100 vol. % oxygen |
Product Description
Controls:
Control principle Continuous flow, pressure-limited,time-cycled
Conventional ventilation IMV/IPPV, CPAP
Triggered ventilation IMV, SIPPV, PSV1), Leak adapted
Trigger Flow/volume trigger, Leak adapted
Trigger delay approx. 30 ms
High Frequency Ventilation1) 3) CPAP+HFV, IMV+HFV
Frequency 5 to 20 Hz
Volume Gurantee Ventilation1) SIMV+VG, SIPPV+VG, PSV+VG
Oxygen mixer loss (bleed flow) 0 (zero) L/min
Monitoring:
Flow Monitoring at the Y-piece, integrated
Volume Monitoring at the Y-piece, integrated
Lung function monitoring Compliance, resistance C20/C, time constant
FiO2 Monitoring integrated
Real-time curves Flow and pressure, integrated
Inspiratory Oxygen concentration 21 -100 vol. % oxygen
Peak Pressure to 99 mbar
Mean Airway Pressure to 99 mbar
Graphic trends 6 parameters, integrated
Logbook record of last 100 alarms
Mechanical Ventilation:
Use of a machine to induce alternating inflation and deflation of the lungs with air
Regulates Oxygen/CO2 levels in blood
Volume and Pressure of Air forced into Lungs is controlled
Has to meet varying demands of patient ventilation
Modes of Ventilation:
Volume control
Pressure control
CPAP/BIPAP
Support Ventilation
Combination Of Modes
Intermittent mandatory Ventilation
Components of Ventilator System:
Compressed Gas Source
Patient Breathing Circuit
Mixer/Blender
Hoses
Optional monitoring
O2
Volume
Classification:
Driving Mechanism
Pneumatic (Mark 7)
Electronic (Siemens 900C)Function
Respirators: Short time ventilator support
Conventional Ventilators :long term, advance Support Portability
Transport/battery backup Patients Adult Pediatric Or Neonatal
Technologies and Jargon:
Turbine technology
High frequency Oscillatory ventilation
Modes IMV, SIMV, PC, VC, S/T etc.
MRI Compatible (Non ferromagnetic components)
Flow Sensors- Volume Measurements
Pressure transducers